I had a couple of problems. The first problem was when I tried to start a new study at the beginning. I got an error saying experiment is active, join another experiment. I joined another experiment and set up a routine proton experiment. I typed in mtune and hit start probe tuning… nothing happened. No lines, nothing came up after hitting autoscale. I restarted VNMRj a few times, logged in/out a few times, and nothing worked.

Yongbo solved the problem by reset acqproc. The issue most likely was due to too many job sent to VnmrJ and caused a dead looping.

Staff Response:

Last Updated April 21, 2015 at 11:45 am by Yongbo